| |
|
|
|
|
|
|
|
|
The
physical principles revolve around the fact that the vast majority
of ultrasound machines detect and calculate the variables of
the Doppler equation, except for the angle between the sound
beam and blood flow that is cos q.
The vascular technologist/sonographer must input this information
into the ultrasound unit via the angle correction control. When
used correctly, the velocity of the blood flow can then be measured.
Usually, the cursor is placed parallel to the walls of the vessel
being investigated to determine angle q.
The ultrasound system software then uses the value of cos q
in the Doppler equation to determine the velocity scale used
to measure the Doppler spectral signals. Since we are dealing
with the cosine of an angle, errors of varying severity are
introduced when an incorrect value is used. When angle correction
is not properly applied, the calculated velocity measurements
may overestimate or underestimate the severity of disease. Utilizing
erroneous velocity measurements to arrive at a final diagnosis
may place the disease severity in a higher or lower category
and thus adversely impact the care the patient receives. Errors are amplified when angles of greater than 60 degrees are used in the determination of velocities. At Doppler angles of less then 60 degrees, an error in the angle correction will result in a very small error in the velocity measurement. As the angle increases between 60 to 90 degrees, the cosine of q (and therefore the denominator of the Doppler equation) rapidly approaches zero; therefore, small errors in the measurement of q result in increasingly large errors in the calculated velocity. It is imperative that a Doppler angle of 60 degrees or less be utilized whenever possible for Doppler velocity measurements. The method of angle correction used must be based on the diagnostic criteria utilized by the laboratory. It is important to use the same techniques as described in the criteria and the same range of angles. A survey of ICAVL accredited laboratories was undertaken and reported by Beatrice Madrazo, MD, RVT, of William Beaumont Hospital in Michigan in 1998. Based on the survey, Dr. Madrazo reported that approximately 72 percent of accredited laboratories were utilizing the diagnostic criteria published by the Eugene Strandness et al from the University of Washington. The guidelines established by the University of Washington for those specific criteria include using a consistent angle of 60 degrees and with the Doppler adjusted parallel to the vessel wall. For all duplex examinations, regardless of the criteria being used, the Doppler angle must be 60 degrees or less. When angles of greater than 60 degrees are obtained, care must be taken to either reposition the transducer or the patient so that the Doppler angle is 60 degrees or less. Occasionally, due to the patient's anatomy, this may not be possible. Whenever angles greater than 60 degrees are used, this must be noted and care must be taken in using these velocity measurements in determining the severity of disease and in reaching a diagnostic conclusion. NEXT>> |
||||||||||||
|
|||||||||||||
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
||
| Copyright 1997-2008 ICAVL, 8830 Stanford Boulevard, Suite 306, Columbia, MD 21045. All rights reserved. | ||||||||